VITROS 350 CHEMISTRY SYSTEM
Report
- Report Number
- 1319681-2012-00230
- Event Type
- Malfunction
- Date Received
- November 30, 2012
- Date of Event
- October 13, 2012
- Report Date
- November 30, 2012
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE INVESTIGATION CONFIRMED THAT A NON-REPRODUCIBLE, LOWER THAN EXPECTED AMON QUALITY CONTROL RESULT WAS OBTAINED ON A VITROS 350 CHEMISTRY SYSTEM. THERE IS NO EVIDENCE THAT THE VITROS AMON REAGENT MALFUNCTIONED. ACCEPTABLE VITROS AMON PERFORMANCE HAS BEEN ACHIEVED USING THE SAME REAGENT LOT FOLLOWING COMPLETION OF "AS-NEEDED" INSTRUMENT MAINTENANCE. THE MOST LIKELY ASSIGNABLE CAUSE OF THE EVENT IS AN INSTRUMENT RELATED ISSUE.
THE CUSTOMER OBTAINED A NON-REPRODUCIBLE, LOWER THAN EXPECTED VITROS AMON QUALITY CONTROL RESULT (QC FLUID D1731= 139.7 VERSUS EXPECTED RESULT= 190.5 UMOL/L) USING A VITROS 350 CHEMISTRY SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THE RESULTS WERE TO RE-OCCUR UNDETECTED USING PATIENT SAMPLES. NO VITROS AMON PATIENT RESULTS RUN DURING THE TIME FRAME OF THIS EVENT WERE QUESTIONED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS, INC. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS 350 CHEMISTRY SYSTEM | CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |