ABBOTT M2000SP
Report
- Report Number
- 3005248192-2012-00019
- Event Type
- Injury
- Date Received
- July 16, 2012
- Date of Event
- June 22, 2012
- Report Date
- June 26, 2012
- Manufacturer
- ABBOTT MOLECULAR, INC.
- Product Code
- JQW
- PMA / PMN Number
- K092705
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
BY FAILING TO WEAR PROTECTIVE EYEWEAR, THE LABORATORY TECHNICIAN DID NOT FOLLOW GOOD LABORATORY PRACTICES AND INSTRUCTIONS PROVIDED IN THE M2000SP OPERATIONS MANUAL FOR EMPTYING THE M2000SP LIQUID WASTE CONTAINER. THE M2000SP OPERATIONS MANUAL 200681-105-OCTOBER 2011 CONTAINS THE FOLLOWING LABELING: USER OR FUNCTION - LIQUID WASTE CONTAINER. CAUTION: THE WASTE IS POTENTIALLY INFECTIOUS. USE APPROPRIATE BIOHAZARD PRECAUTIONS. WASTE HANDLING AND DISPOSAL - LIQUID WASTE. WARNING: THE LIQUID WASTE MAY CONTAIN INFECTIOUS AGENTS AND SHOULD BE CONSIDERED POTENTIALLY INFECTIOUS WASTE. DAILY MAINTENANCE: SERVICE AND MAINTENANCE - EMPTY LIQUID WASTE CONTAINER WARNING: POTENTIAL BIOHAZARD. INSTRUMENT WASTE COULD BE CONTAMINATED WITH POTENTIALLY INFECTIOUS MATERIALS. OBSERVE BASIC BIOHAZARD PRECAUTIONS. WEAR APPROPRIATE PERSONAL PROTECTIVE EQUIPMENT SUCH AS GLOVES, LAB COATS, AND PROTECTIVE EYEWEAR.
A CUSTOMER AT THE (B)(6) HOSPITAL ((B)(6)) REPORTED THAT A LABORATORY TECHNICIAN WAS SPLASHED IN ONE EYE WHEN DISCONNECTING THE LIQUID WASTE CONTAINER ON THE M2000SP INSTRUMENT. THE WASTE CONTAINER WAS BEING EMPTIED AS PART OF THE INSTRUMENT WEEKLY MAINTENANCE. THE SPLASH OCCURRED AS THE WASTE CONTAINER LID WAS BEING REMOVED. THE TECHNICIAN WAS WEARING A LAB COAT AND GLOVES AT THE TIME OF EXPOSURE. ONE OF THE TECHNICIAN'S GLOVES CAUGHT ON THE CONNECTOR ON TOP OF THE LID. LIQUID WASTE WAS FLICKED UP AS THE LID WAS BEING TUNED AND AS THE GLOVE WAS FREED FROM THE CONNECTOR. THE TECHNICIAN'S EYE WAS IMMEDIATELY RINSED. THE TECHNICIAN WENT TO THE HOSPITAL EMERGENCY ROOM TO RECEIVE MEDICAL ATTENTION. ADDITIONAL INFORMATION AS FOLLOWS WAS PROVIDED: THE TECHNICIAN RECEIVED A HBV VACCINATION FOR PROPHYLAXIS; THE TECHNICIAN'S EYE WAS NOT DAMAGED; THE TECHNICIAN ELECTED TO TAKE 1 DOSE OF HIV POST-EXPOSURE PROPHYLAXIS (PEP); HOWEVER SUBSEQUENTLY OPTED TO NOT CONTINUE WITH THE PEP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ABBOTT M2000SP | JQW | ABBOTT MOLECULAR, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |