PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2012-07579
- Event Type
- Injury
- Date Received
- November 30, 2012
- Date of Event
- November 13, 2012
- Report Date
- November 15, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. IT WAS REPORTED A COMMON FEMORAL ARTERY WAS MODERATELY CALCIFIED. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE.
(B)(4). IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS. THE DEVICE WAS REPORTEDLY DISCARDED BY THE HOSPITAL. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A MODERATELY CALCIFIED COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6F SHEATH, AFTER A SUPERFICIAL FEMORAL ARTERY INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. A THIRD PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS.THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20910J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6F |