FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2852711 · Received November 30, 2012

Report

Report Number
1416980-2012-06244
Event Type
Malfunction
Date Received
November 30, 2012
Date of Event
November 16, 2011
Report Date
November 13, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR, INAPPROPRIATE BYPASS OF THE DRAIN, NOT INCLUDING INITIAL DRAIN. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2011, 02:34:04. DURING NIGHT DRAIN CYCLE EIGHT, THE PATIENT'S ULTRAFILTRATION READING WAS 131ML, INDICATING THE HOME PATIENT (HP) DRAINED 131ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 200ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1