ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2012-05398
- Event Type
- Malfunction
- Date Received
- November 30, 2012
- Report Date
- November 7, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED WITH THE BLADE DISCOLORED (BLUE) DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. HOWEVER, THE BLADE OF THE DEVICE WAS NOT BROKEN OFF AS REPORTED. IT IS POSSIBLE THAT THE RETURNED DEVICE WAS THE ONE USED TO COMPLETE THE PROCEDURE AND NOT THE DEVICE COMPLAINED ABOUT. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 INSTRUMENT ERROR SCREEN WAS RECEIVED AND WHEN TESTED WITH GEN04 AN ERROR CODE 5 WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 O INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. THE BLADE WAS FOUND TO BE BROKEN AT THE DISCOLORED (BLUE). PROBABLE CAUSE OF BLADE DISCOLORATION IS APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE ACTIVE BLADE BROKE OFF OUTSIDE OF THE PATIENT. THERE WAS NO OTHER INFORMATION AVAILABLE. THEY USED A SECOND LIKE DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | J9054K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |