FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 2852698 · Received November 30, 2012

Report

Report Number
3005075853-2012-05398
Event Type
Malfunction
Date Received
November 30, 2012
Report Date
November 7, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED WITH THE BLADE DISCOLORED (BLUE) DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. HOWEVER, THE BLADE OF THE DEVICE WAS NOT BROKEN OFF AS REPORTED. IT IS POSSIBLE THAT THE RETURNED DEVICE WAS THE ONE USED TO COMPLETE THE PROCEDURE AND NOT THE DEVICE COMPLAINED ABOUT. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 INSTRUMENT ERROR SCREEN WAS RECEIVED AND WHEN TESTED WITH GEN04 AN ERROR CODE 5 WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 O INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. THE BLADE WAS FOUND TO BE BROKEN AT THE DISCOLORED (BLUE). PROBABLE CAUSE OF BLADE DISCOLORATION IS APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE ACTIVE BLADE BROKE OFF OUTSIDE OF THE PATIENT. THERE WAS NO OTHER INFORMATION AVAILABLE. THEY USED A SECOND LIKE DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA J9054K

Patients

Seq Age Sex Outcome Treatment
1