Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE CUSTOMER REPORTED TO HAVE REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB. COVIDIEN WAS NOT AUTHORIZED TO REPAIR THE VENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) WAS ONLY AUTHORIZED TO UPLOAD THE SOFTWARE. THE VENTILATOR PASSED ALL TESTING.