FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2852611 · Received November 26, 2012

Report

Report Number
3004123209-2012-00512
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
July 15, 2011
Report Date
November 20, 2012
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2010 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. AFTER THIS DATE IT WOULD APPEAR TO HAVE DEVELOPED A FAULT WHERE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. THE DEVICE WAS DISASSEMBLED AND ROUTINE TESTING INDICATED THAT THE Q37 TRANSISTOR WAS FAULTY. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1