FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2852600 · Received November 26, 2012

Report

Report Number
3004123209-2012-00395
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
June 3, 2011
Report Date
November 19, 2012
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2009 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2010. THERE ARE MULTIPLE SELF-TEST FAILS AFTER THIS DATE WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE PAD DEVICE WAS ATTRIBUTED TO A FAULT WITH THE MEMBRANE. THE INITIAL PROBLEM WITH THE DEVICE WENT UNNOTICED BY THE USER AND IT WOULD APPEAR THE DEVICE CONTINUED TO SWITCH ON AUTOMATICALLY WHICH DEPLETED THE PAD-PAK. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE, WHICH IS WHY THE "UNKNOWN" BOX HAS BEEN CHECKED IN THIS REPORT.

Description of Event or Problem · 1

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE DEVICE WOULD NOT POWER-ON. A DEVICE IN THIS FAULT MODE, IF LEFT UNDETECTED COULD RESULT IN THE FAILURE OF THE DEVICE TO PERFORM AS INTENDED IF REQUIRED.