FDA Adverse Event
Malfunction
Summary report: N
HEARTSINE SAMARITAN 300P AND PAD PAK
MDR report key: 2852598
·
Received November 26, 2012
Report
- Report Number
- 3004123209-2012-00391
- Event Type
- Malfunction
- Date Received
- November 26, 2012
- Date of Event
- May 17, 2011
- Report Date
- November 19, 2012
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE PAD DEVICE OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. IT FAILED TO POWER ON AFTER THIS DATE. THE DEVICE WAS DISASSEMBLED AND ROUTINE TESTING TRACED THE PROBLEM TO THE Q55 TRANSISTOR. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
Description of Event or Problem · 1
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE DEVICE WOULD NOT POWER-ON. A DEVICE IN THIS FAULT MODE, IF LEFT UNDETECTED COULD RESULT IN THE FAILURE OF THE DEVICE TO PERFORM AS INTENDED IF REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |