FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2852567 · Received November 26, 2012

Report

Report Number
3004123209-2012-00374
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
March 28, 2011
Report Date
November 16, 2012
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2010 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. SELF TEST FAILURES OCCURRED AFTER THIS DATE AND ALERTING THE USER TO A LOW BATTERY. INDICATOR BEING ILLUMINATED AND AN AUDIBLE BEEP. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT WITH THE POGO-PINS. THE FLUCTUATION IN VOLTAGE WAS CONSIDERED SUFFICIENT TO TRIGGER LOW BATTERY WARNINGS. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.