FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2852560 · Received November 26, 2012

Report

Report Number
3004123209-2012-00366
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
February 16, 2011
Report Date
November 16, 2012
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2010 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2010. THERE ARE MULTIPLE SELF-TEST FAILS AFTER THIS DATE INDICATING THE PAD DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE PAD DEVICE WAS ATTRIBUTED TO A FAULT WITH THE MEMBRANE. ON CLOSER INSPECTION SIGNIFICANT SIGNS OF CORROSION WERE FOUND ON THE MEMBRANE TAIL. THIS WOULD INDICATE THAT THE DEVICE MAY HAVE BEEN STORED IN A LOCATION WHERE IT IS EXPOSED TO ADVERSE ENVIRONMENTAL CONDITIONS. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED COULD RESULT IN THE BATTERY BECOMING DEPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATION MKJ HEARTSINE TECHNOLOGIES, LTD PAD