FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2852476 · Received November 13, 2012

Report

Report Number
1720753-2012-09284
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 18, 2012
Report Date
November 13, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REFORMATTED THE HARD DRIVE AND RELOADED SOFTWARE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THE SYSTEM MIXED UP IMAGES. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1