TERUMO TCM II COLLING AND HEATING SYSTEM
Report
- Report Number
- 1828100-2012-01439
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Date of Event
- October 19, 2012
- Report Date
- October 19, 2012
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REPRESENTATIVE (FSR). THE FSR FOUND THE WATER TO BE EXTREMELY DIRTY. AFTER HE PERFORMED A DE-SCALING, THE FLOWS RETURNED TO NORMAL. THE UNIT WAS TESTED TO MANUFACTURER'S SPECIFICATIONS AND RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THERE WAS NO WATER FLOW IN THE COOL DOWN OR DEFROST MODES. AS A RESULT, AN ALTERNATIVE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WERE NO DELAYS, NO BLOOD LOSS OR NO ADVERSE CONSEQUENCES TO THE PATIENT.