FDA Adverse Event Malfunction Summary report: N

MAGNA-CLAVE STERILIZER

MDR report key: 2852380 · Received November 13, 2012

Report

Report Number
1017522-2012-00026
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 15, 2012
Report Date
October 15, 2012
Manufacturer
PELTON & CRANE
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE MAGNA-CLAVE STERILIZER IS A MANUALLY OPERATED DEVICE REQUIRING THE OPERATOR TO MONITOR THE STERILIZER DURING THE STERILIZING CYCLE INCLUDING CHANGING THE STERILIZER FROM "STERILIZE" TO "VENT" WHEN THE BUZZER ACTIVATES AT THE END OF THE STERILIZATION CYCLE INDICATING THE STERILIZATION CYCLE IS COMPLETE. WHILE THE STERILIZER IS IN THE "STERILIZE" MODE THE "HEAT ON" LIGHT WILL ILLUMINATE INDICATING THE MAIN HEATING ELEMENTS ARE ENERGIZED AS DOCUMENTED IN THE OPERATORS MANUAL. THE MAINTENANCE DEPARTMENT FROM THE USER FACILITY INFORMED PALTON & CRANE THE EVENT OCCURRED AS A RESULT OF USER ERROR BECAUSE THE OPERATOR DID NOT FOLLOW THE MAGNA-CLAVE OPERATING INSTRUCTIONS AS WELL AS ADHERE TO THE "WARNING" ON THE STERILIZER REGARDING THE IMPROPER OPERATION OF THE DEVICE. INSTEAD THE OPERATOR STARTED A STERILIZATION CYCLE AND LEFT THE STERILIZER RUNNING UNATTENDED IN THE "STERILIZE" MODE OVER THE WEEKEND AND DID NOT PROPERLY VENT THE STERILIZER AT THE END OF THE STERILIZATION CYCLE WHICH IN TURN DEACTIVATES THE MAIN HEATING ELEMENTS.

Description of Event or Problem · 1

THE OPERATOR OF A MAGNA-CLAVE STERILIZER STARTED A STERILIZING CYCLE ON FRIDAY EVENING AND LEFT THE STERILIZER RUNNING ON THE "STERILIZE" MODE OVER THE WEEKEND. WHEN THE FACILITY OPENED UP ON MONDAY MORNING THE OPERATOR OPENED THE DOOR TO THE STERILIZER AND SMOKE CAME OUT OF THE STERILIZER CHAMBER AND SET OFF THE SMOKE DETECTOR. THE INSTRUMENT INSIDE THE STERILIZER CHAMBER HAD BURNED. THERE WERE NO INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAGNA-CLAVE STERILIZER STERILIZER FLE PELTON & CRANE A6

Patients

Seq Age Sex Outcome Treatment
1