FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2852174
·
Received November 12, 2012
Report
- Report Number
- 8020893-2012-01204
- Event Type
- Malfunction
- Date Received
- November 12, 2012
- Date of Event
- October 22, 2012
- Report Date
- November 2, 2012
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR STOPPED CYCLING. NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |