FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2852041 · Received November 30, 2012

Report

Report Number
1416980-2012-06221
Event Type
Malfunction
Date Received
November 30, 2012
Date of Event
November 1, 2012
Report Date
November 8, 2012
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED TO BAXTER FOR EVALUATION. THEREFORE, THE REPORTED CONDITION COULD NOT BE CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER UNITED KINGDOM RECEIVED A REPORT THAT AN INFUSOR HAD NO FLOW DURING FILLING. THE DEVICE WAS FILLED WITH A TOTAL VOLUME OF 195 ML OF SOLUTION, 50 MG/ML OF FLUOROURACIL (ACCORD) AND NORMAL SALINE. THIS CONDITION INTERRUPTED THERAPY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12G026

Patients

Seq Age Sex Outcome Treatment
1 50 MG/ML OF FLUOROURACIL (ACCORD)| NORMAL SALINE