FDA Adverse Event Injury Summary report: N

PFNA-II BLADE L90 TAN

MDR report key: 2851946 · Received November 29, 2012

Report

Report Number
2520274-2012-03452
Event Type
Injury
Date Received
November 29, 2012
Date of Event
June 24, 2009
Report Date
July 13, 2009
Manufacturer
SYNTHES
Product Code
HSB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: DEVICE WAS USED FOR TREATMENT. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THIS INDIVIDUAL ADVERSE EVENT REPORT IS BEING MADE IN ACCORDANCE WITH THE SYNTHES MDR EXEMPTION REQUEST DATED (B)(4) 2011. A REVIEW OF THE EVENTS ASSOCIATED WITH THE REQUEST WAS PERFORMED AND IT WAS DETERMINED THAT NONE OF THE EVENTS CONSTITUTES SYSTEMIC ISSUES RELATED TO PRODUCT QUALITY, CHANGES IN PRODUCT DESIGN, METHOD OF USE, OR CHANGES IN EXPECTED RISK THRESHOLDS. REVIEW OF THE DATA ALSO INDICATED NO SIGNIFICANT CHANGE IN RISK/BENEFIT OF EACH PRODUCT CATEGORY, NO EVIDENCE OF DEFICIENCIES IN THE DESIGN, LABELING, OR MANUFACTURE OF THE DEVICE. AS NO LOT NUMBER WAS PROVIDED, NO DEVICE HISTORY RECORD REVIEW CAN BE PERFORMED. THE DEVICE IS NOT BEING RETURNED FOR EVALUATION, NO FURTHER INFORMATION IS AVAILABLE ON THIS EVENT. NO FURTHER INVESTIGATION CAN BE PERFORMED.

Description of Event or Problem · 1

(B)(6) COMPLAINT HANDLING UNIT REPORTED A PROXIMAL FEMORAL NAIL ANTIROTATION-BLADE PENETRATION. ON (B)(6) 2009 A TROCHANTERIC BONE FRACTURE WAS TREATED WITH A (B)(4)PROXIMAL FEMORAL NAIL ANTIROTATION SYSTEM. ON (B)(6) 2009 THERE WAS LITTLE ROTATION OF THE FEMORAL HEAD AND BLADE TELESCOPING. ON (B)(6) 2009, FURTHER ROTATION OF FEMORAL HEAD WAS FOUND. ON (B)(6) 2009, THE PATIENT FELT PAIN AND VISITED THE HOSPITAL. AN OF X-RAY SHOWED BLADE PENETRATION TO THE PELVIC. ON (B)(6) 2009, REVISION SURGERY WAS PERFORMED AND A TOTAL HIP ARTHROPLASTY WAS DONE BECAUSE OF SEVERE PELVIC DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PFNA-II BLADE L90 TAN PFNA-II BLADE HSB SYNTHES

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention