FDA Adverse Event Malfunction Summary report: N

COULTER® LH 500 INSTRUMENT

MDR report key: 2851878 · Received November 29, 2012

Report

Report Number
1061932-2012-02788
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
November 2, 2012
Report Date
November 2, 2012
Manufacturer
BECKMAN COULTER INC.
Product Code
GKZ
PMA / PMN Number
K032000
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2012 FOR THIS EVENT. FOUND A LEAK AT THE PROBE OF THE INSTRUMENT. THE FSE NOTICED THAT THE PROBE WAS BENT AND THE PROBE WIPE RINSE BLOCK WAS MISALIGNED. THE FSE REPLACED THE PROBE AND THE PROBE WIPE RINSE BLOCK AND THE LEAK WAS FIXED. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. AS PER PRODUCT LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A LEAK AT THE PROBE OF THE COULTER LH 500 INSTRUMENT. THE VOLUME OF THE LEAK WAS ABOUT 3 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROPER EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LAB COAT AT THE TIME OF THE OCCURRENCE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. THERE WERE NO ERRONEOUS RESULTS IN CONNECTION WITH THE LEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER® LH 500 INSTRUMENT COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER INC. LH 500 N/A

Patients

Seq Age Sex Outcome Treatment
1