FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 2851809 · Received November 29, 2012

Report

Report Number
2955842-2012-01119
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
October 31, 2012
Report Date
October 31, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE GRIP CABLE IS FRAYED AT THE DISTAL IDLER PULLEY. THE FRAYED STRANDS STICK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. ENGINEERING EVALUATION ALSO FOUND THAT THE FRAYED GRIP CABLE IS DERAILED AT THE DISTAL IDLER PULLEY. THE GRIPS CAN STILL OPEN AND CLOSE; HOWEVER, THE MOVEMENT MAY NOT BE PRECISE. THE CABLE DERAILMENT IS LIKELY DUE TO THE CABLE LOSING CONTACT WITH THE PULLEY DURING WRIST ARTICULATION. THE FLEET ANGLE BETWEEN THE PROXIMAL AND DISTAL PULLEYS MAY CONTRIBUTE TO THE DERAILMENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE MEGASUTURECUT NEEDLE DRIVE INSTRUMENT IT WAS NOTED THAT THE CABLE WAS BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGASUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120809 969

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES.