MEGASUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2012-01119
- Event Type
- Malfunction
- Date Received
- November 29, 2012
- Date of Event
- October 31, 2012
- Report Date
- October 31, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE GRIP CABLE IS FRAYED AT THE DISTAL IDLER PULLEY. THE FRAYED STRANDS STICK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. ENGINEERING EVALUATION ALSO FOUND THAT THE FRAYED GRIP CABLE IS DERAILED AT THE DISTAL IDLER PULLEY. THE GRIPS CAN STILL OPEN AND CLOSE; HOWEVER, THE MOVEMENT MAY NOT BE PRECISE. THE CABLE DERAILMENT IS LIKELY DUE TO THE CABLE LOSING CONTACT WITH THE PULLEY DURING WRIST ARTICULATION. THE FLEET ANGLE BETWEEN THE PROXIMAL AND DISTAL PULLEYS MAY CONTRIBUTE TO THE DERAILMENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING REPROCESSING OF THE MEGASUTURECUT NEEDLE DRIVE INSTRUMENT IT WAS NOTED THAT THE CABLE WAS BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEGASUTURECUT NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10120809 969 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES. |