FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 2851805 · Received November 29, 2012

Report

Report Number
2955842-2012-01114
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
October 26, 2012
Report Date
October 31, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CABLE IS BROKE AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE WIRES ON THE MEGASUTURECUT NEEDLE DRIVER INSTRUMENT FRAYED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGASUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10110625 423

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM AND ACCESSORIES