ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14095
- Event Type
- Malfunction
- Date Received
- November 29, 2012
- Report Date
- November 3, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWS NO EVIDENCE OF A LOAD STEP MALFUNCTION. THE PUMP POWERS ON NORMALLY. AN EZPRIME OPERATION WAS SUCCESSFULLY PERFORMED AND THE PUMP WAS ABLE TO PRIME NORMALLY. FORCE SENSOR READINGS WERE FOUND TO BE WITHIN SPECIFICATIONS. THE PUMP WAS OPENED AND THERE WAS NO DAMAGE FOUND TO THE FORCE SENSOR CIRCUIT OR TO THE POWER CIRCUIT. THERE WAS NO DEFECT FOUND ON INVESTIGATION. THE COMPLAINT COULD NOT BE CONFIRMED OR DUPLICATED.
ON (B)(6) 2012 THE REPORTER CONTACTED ANIMAS TO REPORT AN ISSUE DURING THE LOAD STEP IN WHICH THE PUMP PUSHED SEVERAL DROPS OF INSULIN OUT OF THE FILLED CARTRIDGE. THE PATIENT CONFIRMED THE PUMP HAD NOT BEEN EXPOSED TO TRAUMA OR MOISTURE. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 5 YR |