FDA Adverse Event Malfunction Summary report: N

CUSTOM DEFINED PRODUCT

MDR report key: 2851474 · Received November 29, 2012

Report

Report Number
2015691-2012-18725
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
October 26, 2012
Report Date
November 7, 2012
Manufacturer
EDWARDS LIFESCIENCES DR
Product Code
DXO
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE SINGLE DPT - VAMP PLUS KIT WAS RETURNED FOR EVALUATION. DRY BLOOD WAS VISIBLE AT THE DPT STOPCOCK, SAMPLE SITE AND PRESSURE TUBING CONNECTOR. THE PRESSURE TUBING WAS COMPLETELY DETACHED FROM THE SOLVENT BOND JOINT AT THE DISTAL SAMPLE SITE. INDICATIONS OF BONDING SOLVENT WERE PRESENT ONLY AT 1/3 OF THE BOND AREAS (TUBING AND SAMPLE SITE TUBE HOUSING). THERE WAS ALMOST NO BONDING SOLVENT AT THE OTHER 2/3 OF THE BOND AREAS OR SAMPLE SITE TUBE HOUSING. THE INNER AND OUTER DIAMETER OF THE TUBING WAS MEASURED AND FOUND WITHIN SPECIFICATION. THE COMPLAINT WAS CONFIRMED AND APPEARS TO BE RELATED TO INADEQUATE BONDING AT THE JOINT. VISUAL EXAMINATION WAS PERFORMED UNDER MICROSCOPE AT 10X MAGNIFICATION. AN INVESTIGATION IS OPEN TO DETERMINE THE ROOT CAUSE AND IMPLEMENT NECESSARY CORRECTIVE ACTIONS.

Description of Event or Problem · 1

AS REPORTED, THE PRESSURE TUBING DISCONNECTED FROM THE SAMPLING PORT OF THE VAMP SYSTEM. THE NURSE REPORTED THAT THE BABY "LOST A LOT OF BLOOD"; HOWEVER, NO TREATMENT OR TRANSFUSION WAS NEEDED, AND NO "RELEVANT CONSEQUENCES" WERE NOTED. THERE WAS NO ALARM FOR THE PRESSURE LOSS BECAUSE THE PROBLEM OCCURRED DURING TRANSPORT BETWEEN THE ICU AND TAC; AT THAT TIME THE PATIENT WAS MONITORED ONLY WITH ECG, SPO2 AND NIBP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CUSTOM DEFINED PRODUCT DISPOSABLE PRESSURE TRANSDUCER DXO EDWARDS LIFESCIENCES DR T001691A 59297028

Patients

Seq Age Sex Outcome Treatment
1