INFUSOR
Report
- Report Number
- 1416980-2012-06151
- Event Type
- Malfunction
- Date Received
- November 29, 2012
- Date of Event
- November 6, 2012
- Report Date
- November 7, 2012
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
(B)(4). BAXTER RECEIVED ONE SAMPLE CONTAINING NO FLUID IN THE RESERVOIR. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGN OF PHYSICAL ABNORMALITY. AN ACCURACY FLOW TEST WAS PERFORMED ON THE SAMPLE FOR 43.5 HOURS. AT THE END OF THE FLOW TEST PERIOD, THE INFUSOR PRODUCED THE FOLLOWING FLOW RATES: CALCULATED FLOW RATE = 4.82 ML/HR, NORMALIZED FLOW RATE = 4.94 ML/HR, SPECIFICATION RANGE = 4.50 ? 5.50 ML/HR. THE INFUSOR PRODUCED FUNCTIONAL RESULTS WITHIN THE SPECIFICATION RANGE; THE DEVICE PERFORMED AS EXPECTED. THE REPORTED CONDITION OF AN OVERINFUSION WAS NOT CONFIRMED. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. THIS PRODUCT IS NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510(K) NUMBER.
BAXTER (B)(6) RECEIVED A REPORT OF AN INFUSOR THAT OVERINFUSED DURING PATIENT THERAPY. THE DEVICE WAS FILLED WITH FLUOROURACIL AND SALINE. THE ACTUAL FLOW RATE OF THE INFUSION WAS 230ML/18HR. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 12F026 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FLUOROURACIL |