FDA Adverse Event
Malfunction
Summary report: N
BATTERY 14.8 V, 6.3 AH LI-ION BATTERY
MDR report key: 2851410
·
Received October 30, 2012
Report
- Report Number
- 1218950-2012-03631
- Event Type
- Malfunction
- Date Received
- October 30, 2012
- Report Date
- October 4, 2012
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED THE BATTERY IS NOT CHARGING. THERE WAS NO REPORTED PT INVOLVEMENT. THE PHILIPS FIELD SERVICE ENGINEER CONFIRMED THE BATTERY FAILURE WHERE THE BATTERY WAS NOT PROVIDING ANY POWER AND NOT CHARGING. THE DEVICE DID WORK ON AC POWER. THE BATTERY WAS 2 YEARS OLD. WE WILL CONSIDER THIS A MALFUNCTION OF THE BATTERY WHERE THE BATTERY WAS NOT PROVIDING ANY POWER AND NOT CHARGING.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE BATTERY IS NOT CHARGING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BATTERY 14.8 V, 6.3 AH LI-ION BATTERY | MKJ | PHILIPS MEDICAL SYSTEMS | M3538A | 1021-157-P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |