FDA Adverse Event Malfunction Summary report: N

BATTERY 14.8 V, 6.3 AH LI-ION BATTERY

MDR report key: 2851410 · Received October 30, 2012

Report

Report Number
1218950-2012-03631
Event Type
Malfunction
Date Received
October 30, 2012
Report Date
October 4, 2012
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THE BATTERY IS NOT CHARGING. THERE WAS NO REPORTED PT INVOLVEMENT. THE PHILIPS FIELD SERVICE ENGINEER CONFIRMED THE BATTERY FAILURE WHERE THE BATTERY WAS NOT PROVIDING ANY POWER AND NOT CHARGING. THE DEVICE DID WORK ON AC POWER. THE BATTERY WAS 2 YEARS OLD. WE WILL CONSIDER THIS A MALFUNCTION OF THE BATTERY WHERE THE BATTERY WAS NOT PROVIDING ANY POWER AND NOT CHARGING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE BATTERY IS NOT CHARGING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BATTERY 14.8 V, 6.3 AH LI-ION BATTERY MKJ PHILIPS MEDICAL SYSTEMS M3538A 1021-157-P

Patients

Seq Age Sex Outcome Treatment
1