FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2851393 · Received October 29, 2012

Report

Report Number
1811755-2012-03940
Event Type
Malfunction
Date Received
October 29, 2012
Date of Event
October 1, 2012
Report Date
October 1, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY, WORN BEARINGS AND A WORN ROTOR WERE DISCOVERED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK