FDA Adverse Event Malfunction Summary report: N

ADULT HEATED BREATHING CIRCUIT

MDR report key: 2851364 · Received November 29, 2012

Report

Report Number
9611451-2012-00868
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
November 2, 2012
Report Date
November 2, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K100554
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) THE COMPLAINT RT241 BREATHING CIRCUIT IS IN TRANSIT TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE COMPLAINT DEVICE AND COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT241 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND A HEATER WIRE CONTINUITY TEST WAS PERFORMED WITH A CALIBRATED MULTIMETER. RESULTS: THE COMPLAINT DEVICE DID NOT PASS THE HEATER WIRE CONTINUITY TEST. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 120306. CONCLUSION: THE HEATER WIRE OF THE COMPLAINT DEVICE WAS FOUND TO BE OPEN CIRCUIT. WE WERE UNABLE TO DETERMINE THE CAUSE OF THE OPEN CIRCUIT. RESISTANCE TESTS AND VISUAL INSPECTIONS ARE PERFORMED ON ALL BREATHING CIRCUITS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT AN RT241 ADULT BREATHING CIRCUITED CAUSED AN ALARM ON A MR880 HUMIDIFIER, APPROXIMATELY 30-40 MINUTES AFTER USE. THE ALARM WAS RESOLVED AFTER THE BREATHING CIRCUIT WAS CHANGED. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN THE (B)(6) REPORTED THAT AN RT241 ADULT BREATHING CIRCUITED CAUSED AN ALARM ON A MR880 HUMIDIFIER, APPROXIMATELY 30-40 MINUTES AFTER USE. THE ALARM WAS RESOLVED AFTER THE BREATHING CIRCUIT WAS CHANGED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT HEATED BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT241 120306

Patients

Seq Age Sex Outcome Treatment
1 FISHER & PAYKEL HEALTHCARE MR880 RESP. HUMIDIFIER| FISHER & PAYKEL HEALTHCARE MR880 RESP. HUMIDIFIER