FDA Adverse Event Malfunction Summary report: N

TPS HANDPIECE CORD

MDR report key: 2851289 · Received October 29, 2012

Report

Report Number
1811755-2012-03951
Event Type
Malfunction
Date Received
October 29, 2012
Date of Event
October 18, 2012
Report Date
October 18, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HRX
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CORROSION OF THE CONNECTOR PINS WAS NOTED. THE CABLE WAS NOT RETURNED TO THE USER FACILITY; IT IS NOT REPAIRABLE ONCE IT HAS BEEN DISASSEMBLED.

Description of Event or Problem · 1

A TPS HANDPIECE CORD WAS SENT FOR EVAL BECAUSE IT WAS CAUSING HANDPIECES TO RUN WITHOUT ACTIVATION. THE EVENT OCCURRED DURING A ROUTINE FIELD SVC MAINTENANCE VISIT; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TPS HANDPIECE CORD HRX STRYKER INSTRUMENTS KALAMAZOO 08136

Patients

Seq Age Sex Outcome Treatment
1 UNK