FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2851113 · Received November 29, 2012

Report

Report Number
2531779-2012-14062
Event Type
Malfunction
Date Received
November 29, 2012
Report Date
November 12, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THERE IS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE KEYPAD WAS FOUND TO BE PEELING NEAR THE OK KEYPAD BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. DURING TESTING, ALL KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEYPAD CONTACTS.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2012 STATING THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS WERE STICKING AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE. THE REPORTER DENIED ANY DAMAGE TO THE PUMP OR ANY EVIDENCE OF MOISTURE IN THE PUMP. THE PUMP WAS REPORTEDLY CLEANED WITH A DAMP CLOTH AND THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET. THERE IS NO ADVERSE EVENT WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR