FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2851111 · Received November 29, 2012

Report

Report Number
2024168-2012-07539
Event Type
Injury
Date Received
November 29, 2012
Date of Event
October 22, 2012
Report Date
November 9, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED UNSPECIFIED SUTURE RETRIEVAL ISSUE WAS CONFIRMED AS ANALYSIS OF THE DEVICE DETECTED THAT THE POSTERIOR CUFF TABS REMAINED UNBENT AND UNDAMAGED, WHICH INDICATES THAT THE POSTERIOR NEEDLE DID NOT ENTER AND ADHERE TO THE POSTERIOR CUFF. IN ADDITION, THE POSTERIOR PORTION OF THE FOOT WAS DETACHED AND NOT RETURNED, WHICH SUGGEST THAT UNDUE FORCE WAS USED TO SEVER AND DETACH THE POSTERIOR FOOT PORTION. BASED ON VISUAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER AN INTERVENTIONAL CORONARY PROCEDURE, ARTERIOTOMY CLOSURE WAS ATTEMPTED OF THE COMMON FEMORAL ARTERY USING A PERCLOSE PROGLIDE DEVICE. REPORTEDLY, AN UNSPECIFIED SUTURE RETRIEVAL ISSUE OCCURRED. THE DEVICE WAS REMOVED OVER A GUIDE WIRE AND A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO REPORTED SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THE OPERATOR WAS REPORTED TO BE TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 20615J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6-FRENCH