WOOLY BRUSH CHANNEL BRUSH
Report
- Report Number
- 1528319-2012-00022
- Event Type
- Other
- Date Received
- November 15, 2012
- Date of Event
- September 2, 2011
- Report Date
- November 14, 2012
- Manufacturer
- UNITED STATES ENDOSCOPY GROUP, INC.
- Product Code
- MNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE INVOLVED IN THE REPORTED EVENT WAS DISCARDED AND COULD NOT BE EVALUATED. CLEANING BRUSHES ARE INTENDED TO CLEAN THE INTERIOR LUMENS AND OPENINGS OF A SCOPE TO REMOVE ANY TISSUE, BLOOD OR OTHER DEBRIS REMAINING IN THE SCOPE AFTER USE. FOLLOWING MANUAL CLEANING, THE SCOPE IS TO BE FURTHER CLEANED IN ENZYMATIC CLEANER AND TERMINALLY PROCESSED TO ENSURE THE DEVICE LUMENS ARE CLEAR OF ANY OBSTRUCTIONS. THE USER CLEANING THE ENDOSCOPE SUBJECT OF THIS REPORTED EVENT DID NOT RECOGNIZE THAT A PIECE OF THE BRUSH WAS MISSING AND SUBSEQUENTLY DID NOT TAKE STEPS TO REMOVE THE BROKEN PIECES OF THE CLEANING BRUSH. IT IS IMPORTANT IN THE USE OF CLEANING BRUSHES THAT USERS NOT FORCE THE CLEANING BRUSH THROUGH THE ENDOSCOPE SHOULD THE CHANNELS BE OCCLUDED OR RESISTANCE IS MET.
THE USER FACILITY REPORTED THAT A CLEANING BRUSH HEAD DETACHED DURING MANUAL CLEANING OF AN ENDOSCOPE AND WENT UNDETECTED BY THE USER UNTIL THE PHYSICIAN BEGAN A PATIENT PROCEDURE TO REMOVE A FOOD IMPACTION. THE PHYSICIAN NOTED ONLY THE TIP OF CLEANING BRUSH PROTRUDING FROM THE ENDOSCOPE; THE BRUSH HEAD REMAINED LODGED IN THE SCOPE. THE PHYSICIAN REMOVED THE SCOPE AND COMPLETED THE PROCEDURE SUCCESSFULLY WITH ANOTHER SCOPE; NO PATIENT OR USER INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WOOLY BRUSH CHANNEL BRUSH | MNL | UNITED STATES ENDOSCOPY GROUP, INC. | EA00711612 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |