FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2851057 · Received November 29, 2012

Report

Report Number
1416980-2012-06121
Event Type
Malfunction
Date Received
November 29, 2012
Date of Event
November 18, 2012
Report Date
November 18, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PROBLEM WAS CONFIRMED. THE ROOT CAUSE WAS USE ERROR - OPEN CLAMP. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A SAMPLE EVALUATION WILL NOT BE CONDUCTED. THE LOT NUMBER WAS NOT PROVIDED, THEREFORE NO BATCH REVIEW COULD BE PERFORMED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING SYSTEM ERROR (SE) 2240 ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING DWELL. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE MESSAGE TO HOME PATIENT (HP) AND INFORMED THEM TO START OVER USING NEW SUPPLIES. THE HP SAID THERE WERE OPEN CLAMPS ON UNUSED TUBING. THEY FOLLOWED THE ABOVE AND THE HC WAS OKAY. THE PATIENT WAS CONNECTED EITHER AT THE TIME OF THE ALARM OR OBSERVED AIR. THE PATIENT LINE WAS PROPERLY PRIMED BEFORE CONNECTING. PATIENT EXTENSION LINES WERE NOT USED. THE PATIENT DID NOT DISCONNECT ANY TIME PRIOR TO THE ALARM OR OBSERVED AIR. ALL BAGS WERE PROPERLY CONNECTED. THE SUPPLIES WERE NOT DAMAGED BY AN OUTLET PORT CLAMP OR AN ASSIST DEVICE USED TO MAKE THE CONNECTIONS. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE REPORTER. THE HP WOULD COMPLETE THERAPY WITH NEW SUPPLIES. THE SOLUTION TO THIS ISSUE WAS PROVIDED OVER THE PHONE AND A SWAP OF THE DEVICE WAS NOT NECESSARY. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 47 YR HOME CHOICE