ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14064
- Event Type
- Malfunction
- Date Received
- November 29, 2012
- Report Date
- November 2, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- 080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 09/16/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/19/2013 WITH THE FOLLOWING FINDINGS:A REVIEW OF THE PUMP¿S HISTORY SHOWED NO EVIDENCE OF TIME AND DATE ISSUES. DURING TESTING, THE PUMP POWERED ON NORMALLY WITH VIBRATORY ALERTS AND AUDIBLE SOUNDS. THE PUMP WAS TESTED FOR 24 HOURS WITH NO ERRORS, ALARMS, OR WARNINGS. A 10 UNIT NORMAL BOLUS AND A 10 UNIT AUDIO BOLUS WERE PERFORMED SUCCESSFULLY AND ACCURATELY RECORDED IN THE PUMP¿S HISTORY. THE EZMANAGER MAX 2.0.14 BOLUS DOWNLOAD CORRECTLY MATCHES THE IO SERVICE TOOL DOWNLOAD.
ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS TO REPORT THERE WERE BOLUS DOSES GIVEN BUT NOT FOUND IN THE PUMP HISTORY. THE TECHNICAL SUPPORT REPRESENTATIVE CONTACTED THE PATIENT TO OBTAIN INFORMATION AND TO TROUBLESHOOT THE PUMP; HOWEVER WAS UNABLE TO REACH HIM BY TELEPHONE. THE ISSUE WAS NOT RESOLVED. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | OTP GLUCOSE MGMT SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |