FDA Adverse Event Malfunction Summary report: N

SYNREAM REAMER HEAD Ø12.5

MDR report key: 2850920 · Received November 29, 2012

Report

Report Number
8030965-2012-01408
Event Type
Malfunction
Date Received
November 29, 2012
Report Date
October 31, 2012
Manufacturer
SYNTHES GMBH
Product Code
HTO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE VISIBLE TRACES CAN BE EASILY REMOVED. THEREFORE WE CAN DEFINE, THAT THIS IS AN ACCRETION OF CONTACT CORROSION. ACCORDING OF CORROSION CAN BE GENERALLY SAID, THAT ALL MATERIALS ALSO SO CALLED STAINLESS STELLS ARE CONDITIONALLY RESISTANT TO RUST. THE RUST RESISTANT APPLIES ONLY IF THE MATERIAL IS STORED DRY AND IS FREE OF CONTACT WITH OTHER METAL OBJECTS. ALL METALLIC CONTACTS IN DAMP OR WET CONDITIONS PRODUCE ELECTROLYTIC REACTIONS WITH CONTACT CORROSION AS RESULT. THAT IS NORMAL WEAR AND TEAR. NO PRODUCT FAULT COULD BE DETECTED.

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT. THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM.

Description of Event or Problem · 1

A DEVICE REPORT FROM (B)(6) INDICATED A HOSPITAL IN (B)(6) REPORTED: IT WAS FOUND ALL THEIR SYNREAM REAMER HEADS HAD TRACES OF CORROSION ON THEM. NO FURTHER INFORMATION IS AVAILABLE. THIS IS 11 OF 22 REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNREAM REAMER HEAD Ø12.5 REAMER HEAD HTO SYNTHES GMBH 20761

Patients

Seq Age Sex Outcome Treatment
1