FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2850908 · Received November 29, 2012

Report

Report Number
2024168-2012-07532
Event Type
Injury
Date Received
November 29, 2012
Date of Event
October 6, 2012
Report Date
November 9, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SUTURE PULLING OUT OF THE VESSEL DURING NEEDLE PLUNGER REMOVAL WAS CONFIRMED AS ANALYSIS OF THE DEVICE FOUND THAT A POSTERIOR NEEDLE-TO-CUFF MISS OCCURRED AS INDICATED BY AN UNDAMAGED POSTERIOR CUFF WITH UNDISTURBED TABS; AN INDICATOR THE POSTERIOR NEEDLE HAD NOT ENTERED THE CUFF. THIS WOULD APPEAR TO THE OPERATOR AS IF THE SUTURE HAD PULLED OUT FROM THE VESSEL, WHEN IN FACT THE SUTURE WAS NOT DELIVERED DUE TO THE POSTERIOR NEEDLE-TO-CUFF MISS. BASED ON VISUAL INSPECTION AND FUNCTIONAL TESTING, OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DIAGNOSTIC CORONARY ANGIOGRAM, ARTERIOTOMY CLOSURE WAS ATTEMPTED OF THE COMMON FEMORAL ARTERY USING A PERCLOSE PROGLIDE DEVICE. REPORTEDLY, WHEN THE NEEDLE PLUNGER WAS REMOVED, THE SUTURE PULLED OUT OF THE VESSEL. THE DEVICE WAS REMOVED OVER A GUIDE WIRE AND A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO REPORTED SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THE OPERATOR WAS REPORTED TO BE TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 20615J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6-FRENCH