REPLY
Report
- Report Number
- 1000165971-2012-00457
- Event Type
- Injury
- Date Received
- November 23, 2012
- Date of Event
- November 7, 2012
- Report Date
- November 12, 2012
- Manufacturer
- SORIN CRM S.R.L.
- Product Code
- NVZ
- PMA / PMN Number
- P950029
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS EVENT CONCERNS A DEVICE THAT WAS MANUFACTURED AND USED OUTSIDE THE UNITED STATES. ANALYSIS IS PENDING.
CONNECTION AND ELECTRICAL ISSUES WERE REPORTED RELATIVE TO THE SUBJECT PACEMAKER. DURING THE IMPLANT PROCEDURE, THE SCREWDRIVER TURNED FREELY WHEN TIGHTENING THE VENTRICULAR SET-SCREW. THE SCREWDRIVER WAS TURNED COUNTERCLOCKWISE, THE SET-SCREW WAS TIGHTENED AGAIN AND CLICK SOUND WAS HEARD. THE CONNECTOR PIN WAS PROTRUDED FROM THE CONNECTOR BLOCK AND APPROPRIATE CONNECTION WAS CONFIRMED BY PULL TEST. WHEN THE DEVICE WAS STILL OUTSIDE THE POCKET IT WAS NOTED THAT SOME VENTRICULAR PACING SPIKES WERE MISSING (NOISE SUSPECTED). NORMAL MEASUREMENT VALUES WERE OBTAINED AT POST-IMPLANT CHECK. ON THE SAME DAY VENTRICULAR PACING FAILURE WAS NOTED ON THE ECG MONITOR AT THE HOSPITAL WARD. NO LEAD MIGRATION WAS CONFIRMED BY X-RAY. WHEN THE PACEMAKER WAS CHECKED, PACING AND SENSING FAILURE, LEAD IMPEDANCES ABOVE 3 KOHMS AT THE VENTRICULAR CHANNEL WERE CONFIRMED. A RE-INTERVENTION WAS SUBSEQUENTLY PERFORMED. WHEN THE PHYSICIAN PULLED THE VENTRICULAR LEAD BEFORE UNSCREWING THE SETSCREW, THE LEAD CAME OUT OF THE PORT. THIS ISSUE RECURRED THREE TIMES AFTER RE-INSERTING THE LEAD INTO THE PORT AND TIGHTENING IT WITH CLICK SOUND CONFIRMED. WHEN THE VENTRICULAR LEAD WAS MEASURED BY AN ANALYZER NORMAL VALUES WERE OBTAINED. THE LEAD WAS THEN INSERTED INTO THE ATRIAL PORT WITH AAI SETTINGS AND NO ANOMALIES WERE NOTED, APPROPRIATE VALUES WERE OBTAINED WHEN MEASURED BY THE PROGRAMMER. A NEW PACEMAKER WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPLY | NVZ | SORIN CRM S.R.L. | REPLY DR | 2637 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |