FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2850463 · Received November 26, 2012

Report

Report Number
8020893-2012-01268
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
October 31, 2012
Report Date
October 31, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD AN ERRATIC SCREEN. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND RESEATED THE COLOR I/O PCB. NO PARTS WERE REPLACED. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT 840

Patients

Seq Age Sex Outcome Treatment
1