FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2850463
·
Received November 26, 2012
Report
- Report Number
- 8020893-2012-01268
- Event Type
- Malfunction
- Date Received
- November 26, 2012
- Date of Event
- October 31, 2012
- Report Date
- October 31, 2012
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD AN ERRATIC SCREEN. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND RESEATED THE COLOR I/O PCB. NO PARTS WERE REPLACED. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |