FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2850461 · Received November 26, 2012

Report

Report Number
8020893-2012-01270
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
October 31, 2012
Report Date
October 31, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK GUI DISPLAY WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE CUSTOMER REPORTED TO HAVE REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB. COVIDIEN WAS NOT AUTHORIZED TO REPAIR THE VENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) WAS ONLY AUTHORIZED TO UPLOAD THE SOFTWARE. THE VENTILATOR PASSED ALL TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT 840

Patients

Seq Age Sex Outcome Treatment
1