ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-14050
- Event Type
- Malfunction
- Date Received
- November 29, 2012
- Report Date
- November 12, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4): THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO DAMAGE WAS OBSERVED. DURING TESTING, THE UP ARROW KEYPAD BUTTON WAS FOUND TO BE INTERMITTENTLY UNRESPONSIVE; THE DOWN ARROW, OK AND CONTRAST BUTTONS RESPONDED APPROPRIATELY. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL OF THE BUTTON CONTACTS DURING EVALUATION.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2012 STATING THAT THE UP ARROW KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE. THE REPORTER DENIED ANY DAMAGE TO THE PUMP OR ANY EVIDENCE OF MOISTURE IN THE PUMP. THE PUMP WAS REPORTEDLY NOT CLEANED AND THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET. THERE IS NO ADVERSE EVENT WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |