FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 2850195
·
Received November 21, 2012
Report
- Report Number
- 2518422-2012-02258
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- October 30, 2012
- Report Date
- October 30, 2012
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MFR'S SVC CTR, THE AC INLET CONNECTOR WAS FOUND TO BE DAMAGED. THE VENTILATOR'S AC INLET CONNECTOR WAS REPLACED TO ADDRESS THIS ISSUE.
Description of Event or Problem · 1
THE MFR RECEIVED INFO ALLEGING A VENTILATOR'S POWER PORT (AC INLET) WAS BROKEN. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | 1054260 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |