DUROM ACETABULAR COMPONENT 52/46 CODE L
Report
- Report Number
- 9613350-2012-01097
- Event Type
- Injury
- Date Received
- November 26, 2012
- Date of Event
- September 14, 2012
- Report Date
- October 30, 2012
- Manufacturer
- ZIMMER GMBH
- Product Code
- KWA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHYSICIAN
Narratives
THE MANUFACTURER DID NOT RECEIVE DEVICES OR X-RAYS FOR REVIEW. WHERE LOT NUMBERS WERE RECEIVED FOR THE DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. BASED ON AN EXTENSIVE INVESTIGATION OF EVENTS REPORTED FROM SEVERAL USER FACILITIES OUTSIDE THE US, ZIMMER IDENTIFIED THAT THE MOST PROBABLE CAUSE FOR THE OUTCOME OBSERVED WAS A LOOSE OR UNSTABLE CUP THAT RESULTED FROM USE OF SURGICAL TECHNIQUES NOT CONSISTENT WITH THE MANUFACTURER'S RECOMMENDATIONS. AS A CORRECTIVE ACTION, A RETRAINING PROGRAM FOR USERS OUTSIDE THE US WAS INITIATED IN NOVEMBER 2009 AND REPORTED TO THE NATIONAL COMPETENT AUTHORITIES AS REQUIRED. THE DUROM CUP REPORTED IN THIS CASE IS NOT MARKETED IN THE US. A SIMILAR CUP, COMPATIBLE WITH THE METASUL LDH FEMORAL HEAD, IS CLEARED IN THE US AND A CORRECTIVE ACTION FOR THIS PRODUCT WAS REPORTED TO FDA IN JULY 2008 AS CORRECTION (B)(4). SHOULD ADDITIONAL INFORMATION INCLUDING THE FINAL INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE FILED. THE NEED FOR FURTHER CORRECTIVE MEASURES IS NOT INDICATED AT THIS TIME AND ZIMMER (B)(4) CONSIDERS THIS CASE AS CLOSED. (B)(4).
IT IS REPORTED THAT THE PATIENT RECEIVED AN ACETABULAR CUP AND UNDERWENT A REVISION SURGERY DUE TO PAIN AND HIGH COCR EVALUATED IN LABORATORY RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUROM ACETABULAR COMPONENT 52/46 CODE L | DUROM ACETABULAR COMPONENT AND METASUL | KWA | ZIMMER GMBH | 2363087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Hospitalization| R |