VA-LCP-2-COLUMN DRP2.4 VOLAR R SHAFT 2HO
Report
- Report Number
- 8030965-2012-01394
- Event Type
- Injury
- Date Received
- November 29, 2012
- Date of Event
- October 31, 2012
- Report Date
- October 31, 2012
- Manufacturer
- SYNTHES GMBH
- Product Code
- HRS
- PMA / PMN Number
- K083694
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: DEVICE WAS USED FOR TREATMENT. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE MANUFACTURING RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE HOLE VA2 IS NOT MEASURABLE, BECAUSE THIS HOLE GOT DAMAGED AFTER THE LOT RELEASE. THE OVERALL CONCLUSION TO THE PLATE: ALL INSPECTED CHARACTERISTICS ARE IN SPECIFICATION. WE CONCLUDE THAT THE CAUSE OF THIS DAMAGE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES.
THIS IS 1 OF 2 REPORTS FOR COMPLAINT (B)(4).
DEVICE REPORT FROM (B)(4) REPORTS AN EVENT IN (B)(4) AS FOLLOWS: DURING A DISTAL RADIUS FRACTURE PROCEDURE, AFTER REPOSITIONING AND IMPLANTING THE PLATE, THE SURGEON THEN INSERTED THE CORTEX SCREW IN A POSITIONING HOLE AT THE SHAFT. WHILE THE SURGEON WAS INSERTING THE VA LOCKING SCREW INTO THE MOST DISTAL HOLE IN THE FIRST ROW AT THE ULNAR SIDE OF THE PLATE WITH THE TORQUE LIMITING DRIVER USING A GUIDING BLOCK BY A FIXED-ANGLE PROCEDURE, THE LOCKING SCREW WENT THROUGH THE PLATE. THIS EVENT WAS NOTED WHILE VIEWING THE X-RAY. THE SURGEON REMOVED THE PLATE AND SCREWS AND REPLACED THE PLATE WITH ANOTHER SIZED PLATE. THIS IS 1 OF 2 REPORTS FOR THE SAME EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VA-LCP-2-COLUMN DRP2.4 VOLAR R SHAFT 2HO | PLATE | HRS | SYNTHES GMBH | 7615062 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Required Intervention | SCREW |