STERICHEK TOTAL CHLORINE REAGENT STRIP

MDR report key: 284977 · Received July 7, 2000

Report

Report Number
1833407-2000-00001
Event Type
Other
Date Received
July 7, 2000
Date of Event
June 22, 2000
Report Date
July 7, 2000
Manufacturer
ENVIRONMENTAL TEST SYSTEMS
Product Code
DJW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

A TEST STRIP FOR MEASURING LOW CHLORINE LEVELS IN WATER USED TO PREPARE SOLUTIONS FOR KIDNEY DIALYSIS GAVE FALSELY LOW VALUES IN THE PRESENCE OF VERY HIGH LEVELS OF CHLORINE. THE TESTED WATER CAME FROM A HOLDING TANK THAT HAD BEEN SANITIZED WITH CHLORINE THE NIGHT BEFORE. ON JUNE 27, 2000, A VOICE MAIL REFERRED TO A "PROBLEM" WITH THE NEW REAGENT STRIP FOR LOW RANGE CHLORINE. ON JUNE 28, CO LEARNED THAT TWO PERSONS UNDERGOING KIDNEY DIALYSIS BECAME ILL BECAUSE THE WATER USED TO MAKE THE DIALYSIS SOLUTION CONTAINED CHLORINE. THE DIALYSIS CENTER PERSONNEL CHECKED THE SOLUTION WITH A TEST STRIP DESIGNED TO MEASURE HIGHER LEVELS OF CHLORINE AND SHOWED THAT THERE WAS CHLORINE PRESENT IN THE SOLUTION. THERE WAS INTERVENTION BY DIALYSIS CENTER PERSONNEL AND THE PTS HAVE NOT SUFFERED PERMANENT INJURY. THEY ARE DESCRIBED AS FINE. THE ABOVE INFO IS BASED ON TELEPHONE CONVERSATIONS. THE DIALYSIS CENTER HAS NOT YET PROVIDED A WRITTEN REPORT TO THE MFR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERICHEK TOTAL CHLORINE REAGENT STRIP REAGENT STRIP FOR WATER TESTING DJW ENVIRONMENTAL TEST SYSTEMS * 003040

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other