PULSAR II GENERATOR
Report
- Report Number
- 1226420-2012-00078
- Event Type
- Malfunction
- Date Received
- November 27, 2012
- Date of Event
- October 31, 2012
- Report Date
- January 24, 2013
- Manufacturer
- MEDTRONIC ADVANCED ENERGY, LLC
- Product Code
- MUL
- PMA / PMN Number
- K102029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). METHOD: PRODUCT SCHEDULED FOR RETURN BUT NOT RECEIVED BY MANUFACTURER FOR INSPECTION. RESULTS: PRODUCT SCHEDULED FOR RETURN BUT NOT RECEIVED BY MANUFACTURER FOR INSPECTION. CONCLUSION: PRODUCT SCHEDULED FOR RETURN BUT NOT RECEIVED BY MANUFACTURER FOR INSPECTION PRODUCT EVENT: (B)(4).
PRODUCT EVENT #(B)(4) EVALUATION PROCESS: UNIT RECEIVED IN FAIR CONDITION AND INTERNAL VISUAL INSPECTION REVEALED NO LOOSE OR BROKEN COMPONENTS. TO CHECK FOR POSSIBLE RF LEAKAGE ISSUE STEPS 5.9 AND 5.11 OF THE FINAL TEST (TP-0050) WERE PERFORMED. THE POWER MEASUREMENTS SHOWED THE UNIT CORRECTLY DELIVERING POWER TO RATED LOADS. HOWEVER, IN OPEN-CIRCUIT CONFIGURATION (WHICH REPRESENT THE HANDPIECE NOT IN CONTACT WITH THE PATIENT), THE OUTPUT WAS FOUND TO FLUCTUATE BETWEEN LOW POWER RF LEAKAGE MODE AND NORMAL POWER. THIS WAS FOUND TO BE DUE TO INCORRECT WIRING FOR THE SINGLE CURRENT SENSE TRANSFORMER (T9) ON THE RF BOARD, IT WAS FOUND TO BE BACKWARDS AND WAS CORRECTED. AFTER T9 WIRING CORRECTED THE UNIT RELIABLY WENT INTO RF LEAKAGE MODE DURING THAT PART OF THE TEST. ERROR LOG: 4 E3 (PATIENT RETURN ELECTRODE HAS POOR CONNECTION), 1 E5 (MONOPOLAR HANDPIECE HAS REACHED END OF LIFE), 21 E7 OR E8 (VOLTAGE INTERLOCK ERROR, GENERATOR OVERCURRENT). ERRORS ARE NORMAL USE ERRORS BUT TO ADDRESS OCCURRENCE OF E7 AND E8 470UF CAPACITORS ON DC POWER SUPPLY PCBA WILL BE REPLACED. ROOT CAUSE: THE WEAK POWER OUTPUT COMPLAINT COULD NOT BE REPRODUCED AND THE UNIT WAS FOUND TO DELIVER RATED POWER TO RATED LOADS DURING POWER OUTPUT TESTING. HOWEVER, DURING TEST CASE OF OPEN CIRCUT CONFIURATION THE UNIT WAS FOUND TO EXPERIENCE DIFFICULTY STAYING IN THE REDUCED POWER RF LEAKAGE MODE DUE TO INCORRECT WIRING OF T9 ON RF BOARD. NOTE: PRIOR TO ORIGINAL SHIPMENT THE UNIT IS DOCUMENTED AS HAVING CORRECTLY GONE INTO RF LEAKAGE MODE DURING FINAL TEST. (B)(4).
WEAK POWER OUTPUT OF GENERATOR AND DEVICE.
WEAK POWER OUTPUT OF GENERATOR AND DEVICE.