FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2845305 · Received October 26, 2012

Report

Report Number
1824206-2012-06890
Event Type
Malfunction
Date Received
October 26, 2012
Date of Event
October 4, 2012
Report Date
October 4, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECH REPLACED THE HEAD H-LO DOWN VALVE TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THE HEAD HI-LO DRIFTS DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1