FDA Adverse Event Malfunction Summary report: N

FLEXIVA (TM) TRACTIP HIGH POWER SINGLE-USE LASER FIBER

MDR report key: 2844665 · Received November 26, 2012

Report

Report Number
3005099803-2012-05260
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
November 1, 2012
Report Date
November 1, 2012
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
GEX
PMA / PMN Number
K110685
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT USED PAST EXPIRY DATE.

Additional Manufacturer Narrative · 1

THE FOLLOWING INFORMATION WAS PROVIDED BY THE CUSTOMER ON (B)(6) 2013: THE BLACKENED JACKET WAS AT THE TIP OF THE FIBER DUE TO NORMAL FIBER DEGRADATION AND FUNCTION. THE CUSTOMER ALSO CLARIFIED THAT THE ORIGINALLY REPORTED COMPLAINT WAS REPORTED IN ERROR DUE TO A COMMUNICATION MISUNDERSTANDING. THERE WAS NO FAILURE OF THE DEVICE AND NO COMPLAINT REPORTED ON THIS DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT WITHIN THE FIRST 5 MINUTES OF A URETEROSCOPY PROCEDURE USING A FLEXIVA 200 TRACTIP LASER FIBER, THE BALL TIP OF THE FIBER DEGRADED AT .16 KILOJOULES. WHEN THE ENERGY WAS INCREASED TO .77 KILOJOULES, THE TIP OF THE FIBER; LESS THE BALL WAS PRESENT. HOWEVER, THE JACKET OF THE FIBER WAS BLACKENED APPROXIMATELY 4MM. THE FIBER DID NOT APPEAR TO BE OVERHEATED. THE FIBER WAS POSITIONED 1 MM IN DISTANCE FROM THE KIDNEY STONE. LASER ENERGY FROM A LUMENIS 100 WATT LASER WAS BEING USED WITH THE FIBER. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE WITHOUT ANY COMPLICATIONS TO THE PATIENT WHO IS REPORTEDLY 'FINE' POST PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT WITHIN THE FIRST 5 MINUTES OF A URETEROSCOPY PROCEDURE USING A FLEXIVA 200 TRACTIP LASER FIBER, THE BALL TIP OF THE FIBER DEGRADED AT .16 KILOJOULES. WHEN THE ENERGY WAS INCREASED TO .77 KILOJOULES, THE TIP OF THE FIBER; LESS THE BALL WAS PRESENT. HOWEVER, THE JACKET OF THE FIBER WAS BLACKENED APPROXIMATELY 4MM. THE FIBER DID NOT APPEAR TO BE OVERHEATED. THE FIBER WAS POSITIONED 1 MM IN DISTANCE FROM THE KIDNEY STONE. LASER ENERGY FROM A LUMENIS 100 WATT LASER WAS BEING USED WITH THE FIBER. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE WITHOUT ANY COMPLICATIONS TO THE PATIENT WHO IS REPORTEDLY "FINE" POST PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXIVA (TM) TRACTIP HIGH POWER SINGLE-USE LASER FIBER POWERED LASER SURGICAL INSTRUMENT GEX BOSTON SCIENTIFIC - MARLBOROUGH M0068403960

Patients

Seq Age Sex Outcome Treatment
1 LUMENIS 100 WATT LASER