FDA Adverse Event
Malfunction
Summary report: N
INFINITI VISION SYSTEM OZIL
MDR report key: 2844600
·
Received November 12, 2012
Report
- Report Number
- 2028159-2012-01724
- Event Type
- Malfunction
- Date Received
- November 12, 2012
- Date of Event
- October 11, 2012
- Report Date
- October 12, 2012
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND COULD NOT DUPLICATE THE CUSTOMER REPORTED VACUUM PROBLEM. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. NO SAMPLES WERE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION HAS BEEN PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT VACUUM WAS NOT WORKING DURING A CATARACT EXTRACTION PROCEDURE. THE SYSTEM WAS EXCHANGED. AFTER A 10 MINUTE DELAY, THE PROCEDURE WAS COMPLETED WITH NO HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGEMNTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |