FDA Adverse Event Malfunction Summary report: N

INFINITI VISION SYSTEM OZIL

MDR report key: 2844600 · Received November 12, 2012

Report

Report Number
2028159-2012-01724
Event Type
Malfunction
Date Received
November 12, 2012
Date of Event
October 11, 2012
Report Date
October 12, 2012
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND COULD NOT DUPLICATE THE CUSTOMER REPORTED VACUUM PROBLEM. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. NO SAMPLES WERE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION HAS BEEN PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT VACUUM WAS NOT WORKING DURING A CATARACT EXTRACTION PROCEDURE. THE SYSTEM WAS EXCHANGED. AFTER A 10 MINUTE DELAY, THE PROCEDURE WAS COMPLETED WITH NO HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFINITI VISION SYSTEM OZIL PHACOFRAGEMNTATION SYSTEM HQC ALCON - IRVINE TECHNOLOGY CENTER INFINITI OZIL NA

Patients

Seq Age Sex Outcome Treatment
1