FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2844530 · Received November 5, 2012

Report

Report Number
1720753-2012-09093
Event Type
Malfunction
Date Received
November 5, 2012
Date of Event
October 18, 2012
Report Date
November 5, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE BATTERY CHARGER. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS AN INTERMITTENT PRECHARGE VOLTAGE ERROR. THIS ERROR WILL LIKELY PREVENT THE SYSTEM FROM BOOTING. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1