FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2844434 · Received November 16, 2012

Report

Report Number
1627487-2012-02683
Event Type
Injury
Date Received
November 16, 2012
Date of Event
October 25, 2012
Report Date
October 25, 2012
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTION AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-02682. IT WAS REPORTED THE PT COMPLAINED OF PAIN AT HIS IPG SITE WHILE SITTING FOR LONG PERIODS OF TIME. HE STATED HE DID NOT EXPERIENCE THE PAIN WHILE WALKING OR LAYING DOWN. IT WAS REPORTED THE PHYSICIAN SUSPECTS THE PAIN IS RELATED TO THE IPG PRESSING ON A NERVE. THE PT ALSO STATED HE CHARGES HIS IPG ON A MONTHLY BASIS AND EXPERIENCES POCKET HEATING WHILE CHARGING. THE SJM REPRESENTATIVE ADVISED THE PT OF CHARGING PRECAUTIONS. NO FURTHER INFO IS AVAILABLE AT THIS TIME. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3701 57054

Patients

Seq Age Sex Outcome Treatment
1 31 YR Required Intervention SCS LEAD: MODEL 3228| IMPLANTED: