FDA Adverse Event Malfunction Summary report: N

APEXPRO TELEMETRY SERVER

MDR report key: 2844407 · Received November 5, 2012

Report

Report Number
2124823-2012-00152
Event Type
Malfunction
Date Received
November 5, 2012
Date of Event
October 8, 2012
Report Date
October 8, 2012
Manufacturer
GE HEALTHCARE
Product Code
MHX
PMA / PMN Number
K032369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE SCOPE OF THIS INVESTIGATION IS BASED UPON THE INFO OBTAINED FROM THE FIELD ENGINEER (FE). AFTER ARRIVING ONSITE, THE FE FOUND THAT THE APEXPRO TELEMETRY SERVER (ATS) HAD SHUT DOWN. AFTER POWERING THE UNIT BACK UP, THE FE RAN THE WEBMIN DIAGNOSTICS AND NOTED THAT THE CENTRAL PROCESSING UNIT (CPU) FAN WAS NOT OPERATING. A REPLACEMENT CPU FAN WAS INSTALLED IN THE UNIT ON (B)(4) 2012. THE FE CONFIRMED THAT THE ATS IS NOW BACK ONLINE AND FUNCTIONING NORMALLY AND THE USER IS ABLE TO ADMIT PTS TO THE SYSTEM. REVIEW OF THE SERVICE HISTORY OF THE ATS SHOWED THAT THE CPU FAN HAS NOT BEEN REPLACED SINCE THE UNIT WAS INSTALLED APPROX FOUR YEARS AGO.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE CLINICAL INFO CENTER SHUT DOWN DUE TO AN ISSUE WITH THE APEXPRO TELEMETRY SERVER (ATS) FAN FAILURE. THIS CAUSED LOSS OF MONITORING FOR THIRTY MINUTES AFFECTING FIVE PTS UNTIL ALTERNATE ATS MONITORING WAS ESTABLISHED. THERE WAS NO REPORTED PT INJURY ASSOCIATED WITH THIS EVENT.