FDA Adverse Event
Other
Summary report: N
IMX ANALYZER
MDR report key: 284440
·
Received June 26, 2000
Report
- Report Number
- 1628664-2000-00039
- Event Type
- Other
- Date Received
- June 26, 2000
- Date of Event
- May 19, 2000
- Report Date
- June 21, 2000
- Manufacturer
- ABBOTT LABORATORIES, INC.
- Product Code
- LCI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
CUSTOMER REPORTED IMX BETA-HUMAN CHORIONIC GONADOTROPIN VALUE OF 3158 MIU/ML. VALUE WAS QUESTIONED BY PHYSICIAN. PT WAS REDRAWN AND SAMPLE YIELDED A VALUE OF 21,000 MIU/ML. THE ORIGINAL SAMPLE WAS RETESTED AND YIELDED A VALUE OF 12,884 MIU/ML, CONSISTENT WITH OTHER LAB RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMX ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | LCI | ABBOTT LABORATORIES, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | IMX TOTAL BHCG ASSAY, LIST NUMBER 1A06. |